Chinese Journal of Chromatography ›› 2014, Vol. 32 ›› Issue (5): 477-484.DOI: 10.3724/SP.J.1123.2013.12004

Previous Articles     Next Articles

Screening and confirmation of 24 hormones in cosmetics by ultra high performance liquid chromatography-linear ion trap/orbitrap high resolution mass spectrometry

LI Zhaoyong1,2, WANG Fengmei2, NIU Zengyuan2, LUO Xin2, ZHANG Gang2, CHEN Junhui1   

  1. 1. Research Center for Marine Ecology, The First Institute of Oceanography, State Oceanic Administration, Qingdao 266061, China;
    2. Technical Center for Inspection and Quarantine of Shandong Entry-Exit Inspection and Quarantine Bureau, Qingdao 266002, China
  • Received:2013-12-09 Revised:2014-01-13 Online:2014-05-08 Published:2014-04-26

Abstract:

A method of ultra high performance liquid chromatography-linear ion trap/orbitrap high resolution mass spectrometry (UPLC-LTQ/Orbitrap MS) was established to screen and confirm 24 hormones in cosmetics. Various cosmetic samples were extracted with methanol. The extract was loaded onto a Waters ACQUITY UPLC BEH C18 column (50 mm×2.1 mm, 1.7 μm) using a gradient elution of acetonitrile/water containing 0.1%(v/v) formic acid for the separation. The accurate mass of quasi-molecular ion was acquired by full scanning of electrostatic field orbitrap. The rapid screening was carried out by the accurate mass of quasi-molecular ion. The confirmation analysis for targeted compounds was performed with the retention time and qualitative fragments obtained by data dependent scan mode. Under the optimal conditions, the 24 hormones were routinely detected with mass accuracy error below 3×10-6(3 ppm), and good linearities were obtained in their respective linear ranges with correlation coefficients higher than 0.99. The LODs (S/N=3) of the 24 compounds were ≤10 μg/kg, which can meet the requirements for the actual screening of cosmetic samples. The developed method was applied to screen the hormones in 50 cosmetic samples. The results demonstrate that the method is a useful tool for the rapid screening and identification of the hormones in cosmetics.

Key words: cosmetics, hormones, linear ion trap/orbitrap mass spectrometry (LTQ/Orbitrap MS), ultra high performance liquid chromatography (UPLC)

CLC Number: